User's Manual

3-40 KleeYa - User Manual - Version 2
BASIC FUNCTIONS
OVERVIEW
UNLOADING 1. Pull out the liquid waste container drawer.
2. Notice: Pay attention to the control light on the connector and remove only
unused liquid waste main containers (the control light is off).
Pull the joint level sensor and liquid waste connector (1) out of the full liquid
waste main container (5) and put the liquid waste connector in the tube (2)
behind the containers.
Figure 3-44: Unloading system liquid main container
3. Close the flat cap (3).
4. Close the connector cap (4).
5. Remove the full liquid waste main container (5).
6. Open the liquid waste main container screw cap (6).
7. Empty the liquid waste main container according to the local and national
provisions, legislation and laboratory regulations.
8. Fill preferably 200 ml commercial hypochlorite or bleach into the empty liquid
waste main container.
9. Close the screw cap (6).
LOADING 1. Insert the liquid waste main container (5) into the instrument.
2. Open the connector cap (4).
3. Open the flat cap (3).
4. Plug in the joint level sensor and liquid waste connector (1) into the liquid
waste main container. The tube shall always be guided on the outside of the
liquid waste main container, not in between both containers.