User's Manual

BASIC FUNCTIONS
OVERVIEW
KleeYa - User Manual - Version 2 3-35
MIXTURE Prepare 10 l wash buffer:
1. Open the container screw cap (white cap) outside the instrument.
5x TBS Wash Buffer provided by Diatron is color coded (white lid,
corresponding to white lid on the wash buffer container in the KleeYa
instrument (main and intermediate).
2. If there is still liquid in the container, empty it completely!
3. Open the full 5x TBS wash buffer bottle.
4. Fill the 5x TBS wash buffer completely into the empty wash buffer main
container.
Figure 3-38: Fill in 5x TBS wash buffer
5. Fill the wash buffer main container with 8 l purified water up to the mark.
Figure 3-39: Fill in purified water