User's Manual

BASIC FUNCTIONS
OVERVIEW
KleeYa - User Manual - Version 2 3-29
3.3.4 LOAD OR UNLOAD SYSTEM LIQUID
MAIN CONTAINER
The Liquids detail view shows the liquid level of the wash buffer main container, the
system liquid main container and both liquid waste main containers in percent.
The instrument needs purified water (also called system liquid):
to prime, fill and clean the reagent pipettor probe,
to dilute reagents if requested.
Used system liquid shall comply with the definition of "Instrument Feed Water"
according to CLSI standard for "Preparation and Testing of Reagent Water in the
Clinical Laboratory" (4th Edition), including following characteristics:
pH: 5.0 – 8.0
conductivity: <2 μSi/cm
resistivity: > 0.5 M Ohm*cm
TOC: < 500 ppb
SiO2: <1.0 ppm
Bacteria: <10 CFU/m
Figure 3-31: Liquid detail view
The filling status of the container is additionally displayed in color:
Grey: Liquid level ok.
Orange: Liquid level low.
Red: The container is empty.
Re-usable containers
The system liquid main container is re-usable.