User's Manual

2-4 KleeYa - User Manual - Version 2
INSTRUMENT DESCRIPTION
MODULES
2.2 MODULES
TOUCH SCREEN
AND PC
The system is provided together with the user software, which is located on the PC
of the KleeYa instrument. The user software provides access to the options and
information required to operate the KleeYa instrument and is handled via a touch
screen. All work processes and results are managed by the user software and stored
on the PC.
Optionally, the user software can communicate with devices in the laboratory
network via the PC.
REAGENT
LOADING
The cooled reagent loading bay is used to load Reagent Cartridges via a special rack.
The reagent rack has a RFID label reader on each cartridge position to read
information (e.g. ID, filling level) from the Reagent Cartridges.
SAMPLE LOADING The sample loading bay is used to load test tubes via a special rack. The sample
loading bay has a camera reader to read the information on the 1D or 2D barcode
from the test tubes (samples, calibrators, or controls).
CUVETTE
LOADING
Stacks of cuvettes are inserted via a funnel into the cuvette loader and stored in a
magazine. Single cuvettes are automatically separated and inserted into the
incubator for further processing.
PIPETTORS The instrument has two independent pipettors to pipette reagents, samples,
calibrators or controls into the cuvettes in the incubator. The left pipettor for
reagents has a steel needle. The right pipettor for samples, calibrators or controls
uses AnchorĀ® Tips (disposable tips). The reagent pipettor can be cleaned
automatically with system liquid.
INCUBATOR/
WASHER
The incubator/washer module consists of two rings with 36 cuvette positions each.
The outer incubator ring has two pipetting positions for reagent and sample
pipetting into the cuvette.
The inner washer ring contains three wash stations and allows up to five wash steps
for each cuvette. Magnetic particles are separated on the wall of the cuvette by
magnets located at the wash positions and washed by dispense of wash buffer from
the three wash stations. The inner washer ring contains also further incubation
positions.
Cuvettes are transferred with cuvette pushers between the two rings and into the
measurement unit that is located in the center of the rings.
DISPENSE
CARTRIDGES
The Dispense Cartridges provide the Trigger Solutions (substrates) needed to initiate
the chemiluminescence reaction on the KleeYa instrument. The nature of the trigger
solutions might differ in conjunction with the detection system used by the vendor
of the respective KleeYa assays. The KleeYa instrument provides two possible