User's Manual
INSTRUMENT DESCRIPTION
INSTRUMENT OVERVIEW
KleeYa - User Manual - Version 2 2-1
2 INSTRUMENT DESCRIPTION
2.1 INSTRUMENT OVERVIEW
ENVIRONMENT Compliance with the conditions required in terms of environment and electrical
supply needs to be ensured in order to maintain the performance of the KleeYa
instrument and to guranatee operators safety.
The following environmental conditions are required for operating:
• Ensure a surrounding temperature of 18°C to 30°C (64.4°F to 86°F).
• The Humidity shall be 20 - 80 % non-condensing.
• A clearance of a minimum of 15 cm (5.9 in) from the rear panal to ensure that
the ventilation slits are not blocked to allow heat evacuation.
• Avoid exposure to direct sunlight
INTERLOCK The interlock protective device (i.e cover sensor) of the cover is designed in such a
way that it always gives a stop command to the instrument when the cover is open.
This protective behavior applies even if the interlock protective device is out of
order.
Only the service technician can temporarily disable this protective behavior for
diagnostic, service or maintenance purposes.
UPS The instrument is able to rely on an external Uninterruptable Power Supply (UPS)
interfaced to the instrument PC to prevent susceptibility to power interruption. If
you would like to use an external UPS please contact your distributor for discussion
and setup.
STRATEC is not responsible for an Uninterruptable Power Supply (UPS) used with a
KleeYa System.