MAINTENANCE AND SERVICE
SERVICE
KleeYa - User Manual - Version 2 5-19
5.3.2 AUTOMATIC INSTRUMENT CHECKS/
INITIALIZATION
The instrument performs some checks during power up and initialization automatic
system. For each module, the instrument carried out dedicated sequences for
integrity check.
Those include amongst others:
Measurement unit
Luminometer (start module), lift (moves lift, check drive and light barrier),
Rotor (moves rotor, check drive and light barrier), bubble sensor (start sensor)
Reagent bay / Reagent rack
Bay cooling (checks connection), rack shaker (check shaker movement, check
drive and sensor)
Sample loading
Shutter (closes shutter, check drive and light barrier)
Camera reader for sample recognition
Mirror drive (move mirror, check drive and hall sensor)
Frontcover
LEDs, buzzer, and interlock (initialize), aspirate (start module)
Cabinet
Level monitor (check level sensors, container connection), flooding sensor
(check sensor if it's flooded or not)
Incubator
Incubator (check movement and light barriers)
Disposable Dispense Cartridge Actuator (DDCA)
DDCA (check movement and light barriers)
Washer
Lift (move lift, check drive and light barrier), bubble sensor (start sensor),
valve (check)
Pipettor
Both X-/Y-/Z-drives (check movement and light barrier), reagent and sample
pipettor (check communication)
Cuvette loader
Pusher (move pusher, check drive and light barrier, carousel (move carousel,
check drive and light barrier)