User's Manual
MAINTENANCE AND SERVICE
SERVICE
KleeYa - User Manual - Version 2 5-19
5.3.2 AUTOMATIC INSTRUMENT CHECKS/
INITIALIZATION
The instrument performs some checks during power up and initialization automatic
system. For each module, the instrument carried out dedicated sequences for
integrity check.
Those include amongst others:
• Measurement unit
Luminometer (start module), lift (moves lift, check drive and light barrier),
Rotor (moves rotor, check drive and light barrier), bubble sensor (start sensor)
• Reagent bay / Reagent rack
Bay cooling (checks connection), rack shaker (check shaker movement, check
drive and sensor)
• Sample loading
Shutter (closes shutter, check drive and light barrier)
• Camera reader for sample recognition
Mirror drive (move mirror, check drive and hall sensor)
• Frontcover
LEDs, buzzer, and interlock (initialize), aspirate (start module)
• Cabinet
Level monitor (check level sensors, container connection), flooding sensor
(check sensor if it's flooded or not)
• Incubator
Incubator (check movement and light barriers)
• Disposable Dispense Cartridge Actuator (DDCA)
DDCA (check movement and light barriers)
• Washer
Lift (move lift, check drive and light barrier), bubble sensor (start sensor),
valve (check)
• Pipettor
Both X-/Y-/Z-drives (check movement and light barrier), reagent and sample
pipettor (check communication)
• Cuvette loader
Pusher (move pusher, check drive and light barrier, carousel (move carousel,
check drive and light barrier)