User's Manual
5-14 KleeYa - User Manual - Version 2
MAINTENANCE AND SERVICE
SERVICE
3. Note that the 2D barcode on the label must face up (in the picture on the
right).
Stick the label straight into the recess.
Figure 5-8: Glued label
4. If necessary, add spacers to the sample rack
(see chapter 5.3.1.2 on page 5-14)
.
5.3.1.2 SPACERS FOR SAMPLE RACKS
For 75 mm tubes an additional spacer is required in the sample position of a sample
rack. Please note that all sample positions must be equipped with a spacer.
There are no spacers for 30 mm tubes. These tubes require a transfer tube.
ADD SPACERS Remove the white springs from all 12 sample positions:
1. Use a pen or something similar to loosen the springs at the backside of the
rack.