User's Manual
5-8 KleeYa - User Manual - Version 2
MAINTENANCE AND SERVICE
MAINTENANCE AND CLEANING
5.2.4 QUARTERLY MAINTENANCE
DURATION 2 hrs 12 min (weekly + monthly maintenance)
TOOLS • Soft and lint-free cloths
• Liquinox®
• Distilled water
PROCEDURE 1. Moisten a soft and lint-free cloth with Liquinox® and wipe the steel needle
(SPOLV) of the reagent pipettor.
2. Moisten a soft and lint-free cloth with distilled water and wipe the steel
needle (SPOLV) of the reagent pipettor.
3. Wipe the steel needle (SPOLV) of the reagent pipettor dry with a soft and lint-
free cloth.
4. Fill the additional intermediate container with 4 l sodium hypochlorite (0.5-
0.9%).
5. Pull out the wash buffer/system liquid containers drawer.
6. Place the additional intermediate container to the left of the drawer.
7. Open the cap (1) of the intermediate system liquid container (2) behind the
system liquid container. Make sure that the inner part of the cap (1) with the
connections does not rotate with the outer part.
Figure 5-2: Wash buffer/system liquid containers