User's Manual

MAINTENANCE AND SERVICE
MAINTENANCE AND CLEANING
KleeYa - User Manual - Version 2 5-5
5.2.3 MONTHLY MAINTENANCE
DURATION 1 hrs 30 min (weekly + monthly maintenance), without the duration of the
maintenance task "Database Maintenance"
TOOLS Soft and lint-free cloths
Sodium hypochlorite (0.5-0.9%)
Distilled water
Hypochlorite solution (0.1%)
Additional intermediate container
Clean and empty container
PROCEDURE Surfaces and containers:
1. Wipe the surfaces of the instrument with a soft and lint-free cloth.
2. Pull out the wash buffer/system liquid containers drawer.
3. Clean the system liquid container, system liquid intermediate container, wash
buffer container, and wash buffer intermediate container each with:
a. Fill 1 l sodium hypochlorite (0.5-0.9%) and rinse out.
b. Fill 1 l of distilled water and rinse (repeat 3 times).
4. Close the wash buffer/system liquid containers drawer.
5. Pull out the liquid waste containers drawer.
6. Clean the waste liquid containers and waste liquid intermediate container
each with:
a. Fill 1 l sodium hypochlorite (0.5-0.9%) and rinse out.
b. Fill 1 l distilled water and rinse (repeat 3 times).
c. Fill 1 l hypochlorite solution (0.1%) and rinse out.
7. Close the liquid waste containers drawer.
8. Start the maintenance task "Monthy Maintenance Task"
(see chapter 3.6 on
page 3-75)
.
Washers:
1. Fill the additional intermediate container with 4 l sodium hypochlorite (0.5-
0.9%).
2. Pull out the wash buffer/system liquid containers drawer.
3. Place the additional intermediate container to the left of the drawer.
4. Open the cap (1) of the intermediate wash buffer container (2) behind the
wash buffer container. Make sure that the inner part of the cap (1) with the
connections does not rotate with the outer part.