User's Manual
5-4 KleeYa - User Manual - Version 2
MAINTENANCE AND SERVICE
MAINTENANCE AND CLEANING
5.2 MAINTENANCE AND
CLEANING
5.2.1 DAILY MAINTENANCE
Start of Day
DURATION 6 min
PROCEDURE 1. Start the daily maintenance task "Start of Day Maintenance Task" and fulfill all
instructions
(see chapter 3.6.1 on page 3-77)
.
End of Day
DURATION 4 min
PROCEDURE 1. Start the daily maintenance task "End of Day Maintenance Task" and fulfill all
instructions
(see chapter 3.6.1 on page 3-77)
.
2. Insert the maintenance cartridge on position 1 (left position) and an empty
cartridge on position 2 (right position).
Both cartridges must have a RFID label with a test count of at least 4 to start
the task.
5.2.2 WEEKLY MAINTENANCE
DURATION 10 min
TOOLS • Soft and lint-free cloths
PROCEDURE 1. Wipe the steel needle (SPOLV) of the reagent pipettor with a soft and lint-free
cloth.
2. Wipe the surfaces in the interior of the instrument with a soft and lint-free
cloth.
Do not clean the pipettor guide rods!