USE OF THE INSTRUMENT
UNLOAD AND RELOAD SAMPLES AND REAGENTS
4.8 UNLOAD AND RELOAD
SAMPLES AND REAGENTS
UNLOAD 1. Remove unused/finished sample rack(s) carefully
(see chapter 3.4.1 on page 3-
43)
.
The corresponding LED over the sample loading lane must be off.
2. Remove the unused/finished reagent rack(s)
(see chapter 3.4.2 on page 3-55)
Pay attention to the corresponding state of the reagent loading lane in the
user software. The square above the lane must be gray.
RELOAD
1. Tap on the Load button in the main menu bar to show the Load screen.
2. Tap on the Samples button below the Load button.
3. Load all test tubes into sample rack(s).
4. Load the sample rack(s), one after the other, into the sample loading bay
(see
chapter 3.4.1 on page 3-43)
.
5. If necessary, assign desired tests and dilutions to the samples.
6. Tap on the Reagents button below the Load button.
7. Load all necessary Reagent Cartridges into reagent rack(s).
8. Load the reagent rack(s), one after the other, into the reagent loading bay
(see
chapter 3.4.2 on page 3-55)
.