User's Manual
BASIC FUNCTIONS
DEFINITIONS
KleeYa - User Manual - Version 2 3-89
Start Conditions field:
PROCEDURE Add an assay and specify the conditions:
1. Tap on the Add button.
2. Tap on the Assays tab.
3. Tap on the desired assay in the list.
4. Tap on the Edit tab.
5. Choose the validation method.
6. Specify all parameters of the chosen validation method.
7. Choose one or more start conditions.
Frequency (tests) Specifies after how many tests (same assay) a test with
this control should be performed.
Every (hours) Specifies after how many hours a test with this control
should be performed.
On STAT samples Performs a test with this control if STAT samples are
used.
New reagent cartridge Performs a test with this control when the reagent
cartridge has been replaced.
New calibration curve Performs a test with this control when a new calibration
curve has been recorded.
New ancillary cartridge Performs a test with this control when the ancillary
cartridge has been replaced.