User's Manual

3-88 KleeYa - User Manual - Version 2
BASIC FUNCTIONS
DEFINITIONS
EDIT TAB
Validation Method Allows to select the validation method from the drop-
down menu:
Range: Allows to define the concentration range for
control validation. The control will be only valid if
the measured concentration is within this range.
CV/SD: Allows to define the concentration range for
control validation. The control will be only valid
within the concentration range mean +/- 2 SD.
(E.g. mean=100; 2 SD=10 --> control will be valid if
the concentration is in the range from 90 to 110.)
CV%: Allows to define the Coefficient of Variation in
% for control validation. The control will only be
valid if the CV% of replicates does not exceed this
value.
Cutoff: Allows to select a cutoff value for the control
type. A negative control will only be valid if its
concentration is below the cutoff value. A positive
control will only be valid if its concentration is above
the cutoff value.
Validation method
parameters
Validation method Range (Accuracy Control):
Low (concentration unit): Specifies the lower
concentration range value.
High (concentration unit): Specifies the higher
concentration range value
Validation method CV/SD (Precision Control):
Mean (): Allows to define the expected
concentration of the control.
2 SD (): Allows to define the allowed deviation from
the expected control concentration.
Validation method CV% (Precision Control):
%: Allows to define the maximum CV% between the
replicates of the control.
Validation method Cutoff (Accuracy Control):
Value (): Allows to select the cutoff value for the
control.
Control Type: Allows to select the control type:
positive Pos or negative Neg control.
Replicates Specifies the number of replicates. Some methods
require more than one replicate.