User's Manual
BASIC FUNCTIONS
DEFINITIONS
KleeYa - User Manual - Version 2 3-81
Functions of the Reruns field:
Functions of the Laboratory Normal Result Range field:
Laboratory specific result range: Different to the assay range but within the assay
range set by the assay vendor.
Generals functions:
PROCEDURE
1. Tap on the Definitions button in the main menu bar to show the Definitions
screen.
2. Tap on the Assay button to show the Assay screen.
View/edit an assay:
Editing assays is only possible if no results are available for this assay. Assays can only
be edited if they have never been started or after database maintenance has been
carried out.
1. Tap on the assay selector arrow (next to step 1) to select an assay.
2. View or edit the values of step 1 (General).
Please note the following sub-chapters.
3. Tap on the next steps and view/edit the values.
Allows Reruns Allows reruns of the assay. The rerun rules are set by
the assay developer.
Minimum Lower limit of laboratory normal result range (must be
equal or higher than lower limit of assay range).
Maximum Upper limit of laboratory normal result range (must be
equal or lower than upper limit of assay range).
Delete Assay Allows to delete the selected assay.
Import Allows to import (new) assays.
It is not possible to import an assay that has the same
name as an existing assay. It is possible to give the assay
a new name during import.