BASIC FUNCTIONS
RESULTS
KleeYa - User Manual - Version 2 3-69
Example:
Figure 3-62: Calibration and master curves
A measured calibration curve can be exported and imported as master curve to
transfer the data to different instruments as file. Alternatively, master curves can be
delivered as 2D barcodes on the recalibrators with assay reagent kits (see instruction
for use of the relevant assay for details, indications may vary per assays). A master
curve always needs a recalibration on the instrument in order to create a valid
calibration curve for dose calculation.
Functions:
PROCEDURE
1. Tap on the Results button in the main menu bar to show the Results screen.
2. Calibration or master curves:
• Tap on the Calibration Curves button to show the Calibration Curves
screen.
• Tap on the Master Curves button to show the Master Curves screen.
Allows to configure the table columns
(see
chapter 3.5.1.2 on page 3-62)
.
Import Only for Master Curves!
Allows the import of a master curve from a file.
Delete Only for Master Curves!
Allows the delete a selected master curve.