User's Manual

DOC-06800 Rev. A
6–41
Objet1000 - Connex 2 User Guide
3. confirm this in the wizard screen, In the next wizard screen, observe
the numbers and wait until the level is relatively stabletwo units
above or below the average level shown.
Figure 6-54: Level stabilization
4. Select the Weight is stable check box, and click
Next
.
If you need to calibrate more than one load cell, the next calibration
screen appears. Repeat steps 3 and 4 until all of the load cells are
calibrated.
5. Click
Done
in the final wizard screen.
Figure 6-55: Final wizard screen