User's Manual

DOC-13000 Rev. E 7–49
Connex500/350 User Guide
Testing and
Calibrating the
UV Lamps
Theeffectiveradiationofthe UVlampsusedforcuringmodelscanchange
overtime.Toensureoptimumcuringofmodelsduringprinting,apopup
messageremindsyoutotestthelampsandcalibratetheireffectiveUV
radiationafterevery200hoursofprinting.YoudothisbyrunningtheUV
CalibrationWi
zardandusingthebuiltinUVradiationsensor.
Thewizardcomparestheradiationyoumeasuretotherecommended
radiationlevelforeachofthelamps,ateachprintingmode—HighSpeed,
HighQualityandDigitalMaterial.Theradiationlevelisconsidered
acceptableifitiswithinsevenpercentoftherecommendedlev
el.
Whencalibratingthelamps,thewizardattemptstoadjusttheradiation
level,ifnecessary,andthenpromptsyoutotakeanewreading.Ifthelevel
isacceptable,thewizardcontinuestothenextphase.Iffurtheradjustment
isnecessary,thecurrentphaseisrepeated.Ifthelevelistoolowtobe
adju
sted,thewizardcontinuestothenextphase;thefinalwizardscreen
indicatesthatthelamp’sradiationfortheprintingmodeisunacceptable.
To test and calibrate UV lamp radiation:
1. StarttheUVCalibrationwizardfromtheOptionsmenu(seefigure 726
onpage 23).
2. Makesurethatthebuildtrayisempty,theUVsensorisexposedand
clean,andthattheprintercoverisclosed.Confirmthisinthewizard
screenandclick
Next.
Figure 7-76: Confirmation screen
TheUVlampspowerupandstabilize;thistakesseveralminutes.Then,
theprintblockpassesoverthesensor.
Thewizardcomparesthemeasuredradiationtotherecommended
levelforthelampandprintingmodetested,anddisplaystheresult.
Thewizardcontinuesasfollows:
Ifthelevelmeasuredintheprevioustestisacceptable,thewizard
beginstestingtheleftUVlamp.
Ifthelevelmeasuredintheprevioustestisnotacceptable,the
wizardcalibratestherightUVlampbyadjustingthecurrent
suppliedtoit,thentestsitagain.
Ifthelevelmeasuredintheprevioustestistoolowtobeadjusted,
thewizardrecordsthelamp’sradiationasunacceptable,andbegins
testingtheleftUVlamp.