User manual

Device registers VL6180X
66/79 DocID026171 Rev 6
6.2.47 RESULT__RANGE_REFERENCE_SIGNAL_COUNT
Address: 0x070
Type: R
Reset: 0x0
Description:
6.2.48 RESULT__RANGE_RETURN_AMB_COUNT
Address: 0x074
Type: R
Reset: 0x0
Description:
6.2.49 RESULT__RANGE_REFERENCE_AMB_COUNT
Address: 0x078
Type: R
Reset: 0x0
Description:
313029282726252423222120191817161514131211109876543210
result__range_reference_signal_count
R
[31:0] result__range_reference_signal_count: sensor count output value attributed to signal
correlated to IR emitter on the Reference array.
313029282726252423222120191817161514131211109876543210
result__range_return_amb_count
R
[31:0] result__range_return_amb_count: sensor count output value attributed to uncorrelated ambient
signal on the Return array. Must be multiplied by 6 if used to calculate the ambient to signal
threshold.
313029282726252423222120191817161514131211109876543210
result__range_reference_amb_count
R
[31:0] result__range_reference_amb_count: sensor count output value attributed to uncorrelated
ambient signal on the Reference array.