SR100 Surface Roughness Tester Operating Guide Mode status Display Parameter Status Parameter Selector Mode Selector Start Buttons ArDA Windows Control Unit Traverse Unit ON Button Stylus
Getting Started Storage. The SR100 is supplied with the Traverse Unit (bottom) connected to the Control Unit (top). Before use, separate the Traverse Unit from the Control Unit. Inserting the Batteries Remove battery covers as shown. Then, insert one 3V Lithium 2450 in to the Control Unit and three 3V Lithium 2450s in to the Traverse Unit as shown. Caution: Note: To prevent damage to the stylus, store the SR100 with the units connected. - Use recommended batteries only.
Low Battery Indicators µm Ra µin Rz - Rv Rp Calibrating the Gage µm + Rt Ra Control Unit µin Rz - Rv Rp Traverse Unit Set Up Procedure A new SR100 is fully set up out of the box. However, if the pick-up is changed or another alteration is made, field setup may be required using the following procedure: Press the On button on the Traverse Unit. Position the traverse unit on the calibration standard in the correct orientation.
Specifications Error Codes Traverse length: 0.2" (5mm) E1: Infra Red communication error. Repeat operation. Measurement length: 0.16" (4mm) E2: Measurement attempted when not calibrated. Calibrate the Traverse Unit. Traverse speed: 0.08in/s (2mm/s) Gauge range: 8000µin (200µm) Gauge resolution: 2µin (0.05µm) Display resolution: Ra. 0.4µin (0.01µm) Display resolution: 1µin (0.1µm) for Rz, Rt, Rp, Rv E3: Traverse Unit - Oscillator fault. Retry.