Use and Care Manual

18
Code Description Corrective Action
101
ReceivedSignalToo
Weak/Measuring
TimeTooLong
Usetargetplateorchange
targetsurface.
102
ReceivedSignalToo
High
Targettooreective,usetargetplate
orchangetargetsurface.
201
TooMuchBackground
Light
Reducebackgroundlighton
targetarea.
202
LaserBeam
Interrupted
Removeobstacleandrepeat
measurement.
203
InsufcientBattery Replacebatteries.
301
TemperatureTooHigh Allowdevicetocoolto
operatingtemperature.
302
TemperatureTooLow Allowdevicetowarmupto
operatingtemperature.
401
HardwareError Switchon/offdeviceseveraltimes.
Iferrorcannotbecleared
returndefectiveinstrumentto
ServiceCenterordistributor.
Seewarrantyinformation.
402
UnknownError ContactServiceCenterordistributor.
Seewarrantyinformation.
Error Codes