Product Manual

TECH 500 DP
13
en
7.2 Calibration
The measuring sole calibration is activated with the "CAL"
button.
-CAL- display:
Step 1:
Place the device with the lower measuring sole on as
horizontal a surface as possible (e.g. table) with the
display side facing the user. The first calibration is started
by pressing the
"CAL"
button. CAL flashes in the display.
-CAL2- display:
Calibration step 1 successfully completed.
Step 2:
Turn the unit by 180° in the same position.
Step 3:
The rear of the unit is now facing the user. The second
calibration is started by pressing the
"CAL"
button.
CAL flashes in the display.
rdy display: Calibration completed successfully!
1
2
3
download s
Service
Manual
TECH 500 DP
STABILA
Landauer Str 45
D-76855 Annweiler
Germany
Made in Germany
Tolerance
± 0,05° ± 0,2°
0° /90° 1°...89°
Alkaline
Mignon
www.stabila.com
AA, LR6
Mn 1500