Product Manual
Table Of Contents
- 1. Intended use
- 2. Components of the unit
- 3. Display elements
- 4. Commissioning
- 4.1 Power supply - Charging the Li-ion rechargeable battery
- 4.2 M12 socket pin assignment
- 4.3 Connection cable
- 4.4 Switching the unit on
- 5. Functions
- 5.1 Visual guidance
- 5.2 Acoustic guidance
- 5.3 Automatic display inversion
- 5.4 Setting the "MODE" unit of measurement
- 5.5 Locking the measurement with "HOLD"
- 5.6 Freely selectable zero position "REF"
- 6. "FUNC" button settings
- 7. Checking the measuring tool
- 7.1 Accuracy check
- 7.2 Calibration -– Adjustment
- 7.3 Calibration
- 7.4 Adjusting the sensor
- 7.5 Error messages
- 8. Data transfer
- 8.1 Querying the measurement
- 8.2 Changing the bus address
- 8.3 Error codes
- 8.4 Auto mode
- 8.5 Print mode
- 9. STABILA Analytics evaluation software (optional)
- 10. Technical data
TECH 1000 DP
19
en
8.
Data transfer
Multiple-participant mode:
This mode enables multiple participants with different unit addresses but
the same baud rate to connect to the MODBUS.
MODBUS function Description
Query current angle in 1/100°
Query print angle in 1/100°
Software version
Serial number 1
Serial number 2
Change bus address
Switch off measuring unit
Analysis of data connection
protocol
The transmission protocol is configured in accordance with the MODBUS standard.
Characterformat: 1 start bit, 8 data bits, 2 stop bits, no parity
Baud rate:
Default setting: 9600 Bd
Possible: 1200 Bd ... 19,200 Bd
Idle period: at least 3.5 characters between two messages
Unit address:
Default setting: 032
d
Possible: 001
d
….. 247
d
Please note: The data connection is interrupted if there is no query for > 2 seconds.
To prevent errors when analysing measurements, the angle is only transmitted in degrees.
The REF and HOLD settings are deleted.