TECH 1000 DP
16
en
1
2
7.4 Adjusting the sensor
Step 1:
Once "Sensor adjustment" has been selected with the
"CAL/ADJ" button, confirm by pressing "Enter".
Display: ADJ4P
Step 2:
Hold the unit in plane 1.
Press the "CAL/ADJ" buttons.
If the plane has been adjusted successfully, it is displayed
permanently.
Step 3:
Turn the unit by 90° to plane 2.
Press the "CAL/ADJ" button.
If the plane has been adjusted successfully, it is displayed
permanently.
3
2
1