Product Manual
Table Of Contents
- 1. Intended use
- 2. Components of the unit
- 3. Display elements
- 4. Commissioning
- 4.1 Power supply - Charging the Li-ion rechargeable battery
- 4.2 M12 socket pin assignment
- 4.3 Connection cable
- 4.4 Switching the unit on
- 5. Functions
- 5.1 Visual guidance
- 5.2 Acoustic guidance
- 5.3 Automatic display inversion
- 5.4 Setting the "MODE" unit of measurement
- 5.5 Locking the measurement with "HOLD"
- 5.6 Freely selectable zero position "REF"
- 6. "FUNC" button settings
- 7. Checking the measuring tool
- 7.1 Accuracy check
- 7.2 Calibration -– Adjustment
- 7.3 Calibration
- 7.4 Adjusting the sensor
- 7.5 Error messages
- 8. Data transfer
- 8.1 Querying the measurement
- 8.2 Changing the bus address
- 8.3 Error codes
- 8.4 Auto mode
- 8.5 Print mode
- 9. STABILA Analytics evaluation software (optional)
- 10. Technical data
TECH 1000 DP
15
en
B
C
D
7.4 Adjusting the sensor
The sensor must be adjusted if the "temperature" or "Adj."
symbols are shown in the display.
A:
All 4 planes are adjusted during the sensor adjustment.
B:
The sensor can only be adjusted if the two black bars appear
on the display (in the range of 0°, 90°, 180° and 270°).
C:
"ADJ" flashes during sensor adjustment of the respective
plane.
D:
Planes that have not been adjusted are not displayed.
Successfully adjusted planes are permanently indicated in
the display.
A
2
1
3
Adj.