7.1 Accuracy check
To prevent measuring errors, the accuracy of the
measuring tool must be checked at regular intervals;
for example, each time before beginning work,
or after a heavy impact or extreme fluctuations in
temperature.
Step 1:
Place the unit with the lower measuring sole on as horizon-
tal a surface as possible (e.g. a table) with the display side
facing the user. Determine the measurement.
Step 2:
Turn the unit by 180° in the same position.
Step 3:
The rear of the unit is now facing the user.
Calculating the error:
0.00° − display
= A
360.00° − display
= B
If A+B is greater than ± 0.05°, the tool must be recalibrated.
7.2 Calibration -– Adjustment
By pressing the "CAL/ADJ" button repeatedly, the user can
switch between CAL2P = calibration in relation to the meas-
uring sole and ADJ4P = sensor adjustment. The selected
function is confirmed by pressing the "ENTER" button.