User's Manual
BODYGUARDIAN CONTROL UNIT – BASE KIT Operator Manual
37
`
Led states
OFF
Led Off
SLOW
Led Turned on every 10 Sec & OnTime 1000 ms
MEDIUM
Led Turned on every 1 Sec & OnTime 500 ms
FAST
Led Turned on every 0.5 Sec & OnTime 100 ms
ON
Led On
d.n.c.
Do Not Care
LED Priority
High
H
Medium
M
Low
L
HW/SW driven
SW driven
SW driven
LED1
LED2
LED3
Charging Led
Operative Led
Event Led
Yellow
Green
Yellow
STATE
STATE
Priority
STATE
Priority
Device
BGD OFF
OFF
OFF
OFF
ON charge
ON (by HW)
d.n.c.
d.n.c.
FULL charge
OFF
d.n.c.
d.n.c.
Identification
d.n.c.
Medium-Q
H
Medium-QN
H
Modes
IDLE
d.n.c.
ON
L
OFF
ENGAGED
d.n.c.
L
d.n.c.
SERVICE
ON (by SW)
Medium-Q
H
Medium-QN
H
---> SERVICE (FOTA)
ON (by HW)
d.n.c.
d.n.c.
---> SERVICE (ACCEPT-
TEST)
Test driven
MONITORING
d.n.c.
SLOW
L
d.n.c.
STREAMING
d.n.c.
FAST
M
d.n.c.
Events
Symptomatic Event
Not considered
Electrodes Detached
d.n.c.
d.n.c.
FAST
M
Low Battery Level
SLOW (by SW)
d.n.c.
d.n.c.
Communic.
Active Connection
(In/Out)
d.n.c.
FAST
M
d.n.c.
Data Transfer Failure
d.n.c.
d.n.c.
ON
L