Table 3 - Electromagnetic Immunity for non-life supporting equipment
Immunity
Test
IEC 60601-1-2 Test
Level
Compliance
Electromagnetic
environment
Portable and mobile RF
communications equipment
should be used no closer to
any part of this medical
device, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter
(table 4 - Recommended
separation distance).
Radiated RF
fields
IEC 61000-4-
3
3 V/m
from 80 MHz to 2.5
GHz
IEC 60601-1-
2
Test Levels
3 V/m
d = 1.2 P 80 MHz a 800
MHz
d = 2.3 P 800 MHz a 2.5
GHz
Radiated RF
fields
IEC 61000-4-
6
3 V
from 150 kHz to 80
MHz
IEC 60601-1-
2
Test Levels
3 V
d = 1.2 P
Where P is the maximum
output rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths for fixed RF
transmitter, as determined by
an electromagnetic site
survey, should be less then
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol: