User's Manual
BODYGUARDIAN CONTROL UNIT – BASE KIT Operator Manual
24
Table 2 - Electromagnetic Immunity
Immunity
Test
IEC 60601-1-2 Test
Level
Compliance
Electromagnetic
environment
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
6 kV in contact
8 kV on air
IEC 60601-1-2
Test Levels
Residential/Hospital
(Note 1)
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power supply
line
1 kV for input/output
lines >3m
IEC 60601-1-2
Test Levels
Residential/Hospital
(Note 2)
(Note 3)
Surge
IEC 61000-4-5
1/0.5 kv differential
mode
2/1/0.5 kV common
mode
IEC 60601-1-2
Test Levels
Residential/Hospital
(Note 2)
(Note 3)
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0 % of rated voltage
(voltage dip 100 %)
for 0.5 cycles
40 % of rated voltage
(voltage dip 60 %)
for 5 cycles
70 % of rated voltage
(voltage dip 30 %)
for 25 cycles
0 % of rated voltage
(voltage dip 100 %)
for 5 seconds