User's Manual Part 2

Tech-X Flex User Guide - Firmware v06.50 Tech-X Flex
®
(NG2)
8-26
Intro
Overview
Wi-Fi
Ethernet
System
IP/Video
MoCA
RF
Specs
Supported threshold ranges
8.8.3 RF Script settings
(Select System > System/Module Settings > RF Video Module > RF Script Settings)
The settings in this area apply to the Close-Out Test Script (see Close-Out Test Script on page 8-10)
and include:
Table 8-2 General tab
Threshold Range
Max Digital Power
Min Digital Power
-30 to 30 dBmV
Min 64-QAM MER 24 to 29 dB
Min 256-QAM MER 28 to 33 dB
Max Pre-FEC BER
Max Post-FEC BER
0.0000000010 to 0.001
Max Analog Video Power
Min Analog Video Power
-30 to 30 dBmV
Minimum and maximum OOB channel
power levels
-35 to 35 dBmV
Setting Description
Close-Out
Saved
Results
Format
Controls the format of the results file to be transferred from the
unit, as follows:
XML-1 - Produces a results file in XML format that adheres to a
Spirent-provided schema.
TXT - Produces a results file in a general text format without
XML markup.
Close-Out
Test
Channels
Channels to test during each iteration of the script. Up to five
channels may be specified and each channel must match a
channel configured in the active channel guide, similar to the
normal Channel Sweep test (see Lineup tab on page 8-8 and
Download RF Channel Guide(s) on page 8-23).