User's Manual

27 28
Pop-up Messages (Cont.) Condition Buttons Resolution
Please specify a Stimulus
Response before programming
Attempt to program a set of
pulse parameters without a
Stimulus Response Name.
“OK” Select stimulus
response.
Programmer battery is low.
Please recharge.
PDA battery reaches 30%. “OK” Recharge as soon as
possible.
Programmer is booting after
a reset. When you click OK
the Programmer will switch
off. Press the Power button to
restart.
Hardware reset Button
on the PDA was pressed,
Programmer detected an
error or the Programmer is
launched for the rst time.
Press Power button.
Programmer storage space is
low. Please contact your Spinal
Modulation representative for
maintenance.
PDA storage is at full
capacity (log les are stored
on the PDA with each
signicant operation such
as programming). Message
is displayed and le logging
continues by deleting old
log les to make room
for new log les. Normal
operation continues after user
acknowledgement.
“OK” Contact your
Spinal Modulation
Representative.
Stimulation for one or more
leads has been turned OFF.
One or more leads turned off
due to Low Impedance.
“OK” Perform impedance
measurement and re-
enable if within range.
Otherwise, investigate
lead integrity.
Stimulation has been turned
OFF due to a magnet.
Stimulation can be turned
off by applying a magnet for
the duration specied by the
Magnet Stim On/Off Time
programmable parameter.
“OK” Stimulation can be
turned back on by
applying magnet for
the same “Magnet
Stim On/Off Time”.
Stimulation has been turned
OFF.
Stimulation can be turned off,
either by the “All Stim OFF”
software button on the INS
programmer, or the Off switch
on the TNS.
“OK”
Stimulator battery has reached
End of Service (EOS).
Stimulation has been turned
OFF permanently.
Battery has reached EOS
voltage. Stimulation is
disabled in order to preserve
power for RF communication.
“OK” Schedule replacement
of the Stimulator.
Pop-up Messages (Cont.) Condition Buttons Resolution
Stimulator battery has reached
the Elective Replacement
Indicator (ERI).
Battery has reached ERI
voltage.
“OK” Schedule replacement
of the Stimulator.
Stimulator clock is HH:MM. Do
you want to update Stimulator
clock?
The NS internal time is out
of sync with the Programmer
time.
“Yes”
“No”
The stimulator has been set to
default values.
The Programmer has
encountered an NS with
unreadable or invalid data.
Parameters have been set and
programmed to default values.
“OK” Setup device
parameters as
desired.
The stimulator has unreadable
data. Please reconnect with
programmer in FCE mode or
contact your Spinal Modulation
representative
NS has unreadable data (such
as Trim data)
“OK” Contact your
Spinal Modulation
Representative.
The stimulator is in Upgrade
Mode. Please reconnect with
programmer in FCE mode or
contact your Spinal Modulation
representative.
NS in Boot mode. All
stimulation is disabled.
“OK” Contact your
Spinal Modulation
Representative.
Unable to connect to
stimulator. Make sure the
programmer is close to the
stimulator and try again.
Can’t connect to the NS. “OK” Move the programmer
above the Stimulator
in circular motions.
Error messages may contain additional troubleshooting information such as “
Error Code: 04-123,
BadParameterValue
”. This is an aid for Spinal Modulation engineers to debug errors.