User's Manual
38
Programmerstoragespaceislow.
PleasecontactyourSpinalModulation
representativeformaintenance.
PDAstorageisnearingfullcapacity
(logfilesarestoredonthePDAwith
eachsignificantoperationsuchas
programming).Messageisdisplayed
andfileloggingshouldcontinue.
Normaloperationcontinuesafter
useracknowledgement.
“OK” Contactyour
Spinal
Modulation
Representative.
Stimulationforoneormoreleadshas
beenturnedOFF.
Oneormoreleadsturnedoffdueto
LowImpedance.
“OK” Performimpedance
measurementandre‐
enableifwithinrange.
Otherwise,investigate
leadintegrity.
StimulationhasbeenturnedOFFdueto
amagnet.
Stimulationcanbeturnedoffby
applyingamagnetfortheduration
specifiedbytheMagnetTurnoffTime
programmableparameter.
“OK”
StimulationhasbeenturnedOFF. Stimulationcanbeturnedoff,either
bythe“AllStimOFF”softwarebutton
ontheINSprogrammer,ortheOff
switchontheTNS.
“OK”
StimulatorbatteryhasreachedEndof
Service(EOS).Stimulationhasbeen
turnedOFFpermanently.
BatteryhasreachedEOSvoltage.
Stimulationisdisabledinorderto
preservepowerforRF
communication.
“OK” Schedulereplacement
oftheStimulator.
Stimulatorbatteryhasreachedthe
ElectiveReplacementIndicator(ERI).
BatteryhasreachedERIvoltage. “OK” Schedulereplacement
oftheStimulator.
StimulatorclockisHH:MM.Doyou
wanttoupdateStimulatorclock?
TheNSinternaltimeisoutofsync
withtheProgrammertime.
“Yes”
“No”
Thestimulatorhasbeensettodefault
values.
TheProgrammerhasencounteredan
NSwithunreadableorinvaliddata.
Parametershavebeensetand
programmedtodefaultvalues.
“OK” Setupdevice
parametersasdesired.
Thestimulatorhasunreadabledata.
Pleasereconnectwithprogrammerin
FCEmodeorcontactyourSpinal
Modulationrepresentative
NShasunreadabledata(suchasTrim
data)
“OK” ContactyourSpinal
Modulation
Representative.
ThestimulatorisinUpgradeMode.
Pleasereconnectwithprogrammerin
FCEmodeorcontactyourSpinal
Modulationrepresentative.
NSinBootmode. “OK” ContactyourSpinal
Modulation
Representative.
Unabletoconnecttostimulator.Make
suretheprogrammerisclosetothe
stimulatorandtryagain.
Can’tconnecttotheNS. “OK” Movetheprogrammer
abovetheStimulatorin
circularmotions.
Error messages may contain additional troubleshooting information such as “Error Code:04‐
123,BadParameterValue”.ThisisanaidforSpinalModulationengineerstodebugerrors.