User's Manual

37
AppendixIII:Troubleshooting
PopupMessages Condition Buttons Resolution
AllstimulationhasbeenturnedOFF. Allstimulationturnedoffduetodata
corruptionorDevicebeingin
Hibernatemode.
“OK” ContactyourSpinal
Modulation
Representative.
CancelledConnectRequest Cancelwaspressedwithconnecting
attemptinprogress.

Changessincelastprogrammingwere
lostduetolossofconnectionwiththe
stimulator.Pleasereconnectto
stimulator.
Communicationwaslostpriorto
programmingattempt.
“OK” Reconnectto
stimulatorandreenter
programchanges.
Connectedtostimulator.
Communicationispoor.
AllRFchannelshavenoiselevels
abovethenoisethreshold.
 Trymovingtoanarea
withfewersourcesof
noise.
Connectingtostimulator...“Connect”buttononthemainmenu
waspressed.

ConnectionEstablished. Connectionattemptwassuccessful.“OK”
Connectionwithstimulatorwaslost.
Pleasereconnect.
DroppedRFconnection. "Yes"
"No"
"Cancel"
Doyouwanttoprogramchanges? AnewWorkspaceorExitbuttonwas
selectedwithoutsaving
(programming)changes.
“OK”
InvalidFCEpassword.Pleasetryagain. InvalidFCEpasswordwasentered. “OK” OnlySpinalModulation
representativesshould
accesstheFCEmode.
LeadNdetectedaCurrentTooHigh
condition.
Duringanimpedancemeasurement,
themeasuredcurrentwastoohigh.
“OK” Repeatmeasurement.
Ifproblemreoccurs,
contactyourSpinal
Modulation
representative.
LeadNimpedanceofNNNΩisoutof
range.
Leadimpedanceisoutofrange“OK” Repeatmeasurement
oracceptasis.
Maximumstimulationoutputhasbeen
reached.
Maximumstimulationoutputhas
beenreached(5.2V).
“OK” Investigatelead
integrity.
PleasespecifyaStimulusResponse
beforeprogramming
Attempttoprogramasetofpulse
parameterswithoutaStimulus
ResponseName.
“OK” Selectstimulus
response.
Programmerbatteryislow.Please
recharge.
PDAbatteryreaches30%.“OK” Rechargeassoonas
possible.
Programmerisbootingafterareset.
WhenyouclickOKtheProgrammerwill
switchoff.PressthePowerbuttonto
restart.
HardwareresetButtononthePDA
waspressedortheProgrammeris
launchedforthefirsttime.
PressPowerbutton.
Programmerstoragespaceisfull.File
logginghasstopped.Pleasecontact
yourSpinalModulationrepresentative
formaintenance.
PDAstorageisfull.Afteruser
acknowledgement,normaloperation
continues.
“OK” ContactyourSpinal
Modulation
Representative.

PopupMessages(Cont.) Condition Buttons Resolution