User's Manual

28
AppendixII:INSBatteryLife
ThefollowingfourtablesestimateINSbatterylifeunderthegivenloadimpedanceconditions.
Tables 1 through 4 list estimated longevity based on 1lead stimulation, active 24 hours a day. To
estimatelongevityformultipleactiveleadsandtoaccountfordailyusage,usethefollowingformula:
Where,
C2 =
Longevity for Lead 2 Settings
Daily Usage Lead 2
-35
3014
And,
DailyUsageLead1,2,3and4arefractionalvaluesequaltothenumberofhoursstimulationisenabled
foreachleaddailydividedby24hours.
Note:DonotentervaluesforC1,C2,C3orC4forcorrespondingleadsthatareinactive.
Example:
Estimate longevity for 2 lead stimulation with lead 1 settings of 0.8mA, 40Hz, 200µs across 600 ohms
continuously stimulating, and lead 2 settings of 1.0mA, 60Hz, 400µs across 1Kohm stimulating for 16
hoursperday.
FromTable1(600ohms),longevityforlead1settings=58.9months
FromTable
2(1Kohm),longevityforlead2settings=34.1mon ths
CalculatedC1=(3014/58.935)*(24/24)=16.2
CalculatedC2=(3014/34.135)*(16/24)=35.6
Estimatedlongevityforthisexample=3014/(16.2+35.6
+35)=34.7months