User's Manual
20
• Location:Enterthespinallevelwherestimulationtherapyisdeliveredbythistriallead
• Pulse Parameters: To select and change pulse parameters, first press the desired increment
level:Fine(>),Medium(>>),Coarse(>>>).
Amplitudesbelow2.0mA(>:25µA,>>:50µA,>>>:200µA)
Amplitudesabove2.0mA(>:50µA,>>:100µA,>>>:400µA)
PulseWidth(>:10µs,>>:40µs,>>>:100µs)
Frequency(>:2Hz,>>:4Hz,>>>:10Hz)
TheUP(^)andDown(∨)buttonsnexttothespecific pulseparameterwill
allowtheuser
tochangethesettingatthedesiredincrements.
Thefollowingtableliststhepulseparameters,theirrange,incrementsanddefaultvalue:
Specifications Range StepSize DefaultValue
PulseAmplitude‐PA(µA)
(Dependingonmeasuredimpedance)
0–6000µA 25µA:0‐2000µA
50µA:2000‐6000µA
0µA
MaximumPulseAmplitude‐Max(µA)
ProgrammablebyPatient
SameasPA SameasPA 0µA
PulseWidth–PW(µs) 40–720µs 10µs 40µs
PulseFrequency‐PF(Hz) 4–100Hz 2Hz 60Hz
• Maximum Amplitude:Enter the maximum stimulation amplitude, from the clinically set
amplitudeupto6.0mAthatthepatientisallowedtosetforeachlead.
WARNING: Unlessthestimulationsettingsareknownforaspecificpatient,startwitha
PulseAmplitudeof0µA.
• StimulusResponse:Allowsyoutoassignadescriptortoasetofprogrammedpulseparameters.
Thedescriptoriscomposedofa bodyregionwherethesensationisfeltandadescriptionofthe
sensation.(Eg.UpperBack& Massaging Æ Upper BackMassaging). A StimulusResponsemust
beselectedin
ordertoProgramthesetofpulseparameters.TheLoadOtherbuttonpullsdown
a drop down menu and allows the user to load another Stimulus Response that has been
previouslysavedforthatlead.
NOTE:When restarting stimulation, increase the amplitude slowly until the desired effect is
achieved.