User's Manual

19
tousethisInstantImpedancevalue fortherapydelivery,pressthe“TransferInstantImpedance”
button(“Æ“).
NOTE:The patient may feel the effect of the impedance measurement.Alert the patient to the
possiblestimulation.
NOTE: A transferred impedance value is required before other stimulation parameters can be
selected.
Enable:
Select “ON” to enablethe lead so that itprovides stimulation therapy to the patient.
Select“OFF”iftheleadisnotbeingused.
o WhenEnableisON,the“ON”buttonwillturnthecolorgreen.
o WhenEnableisOFF,the“OFF”buttonwillturnthecolorblack.
o Thebuttonborderisrediftheactivationstateisdifferentfromtheprogrammedvalue.
WARNING:Once Enable is ON for this target, any parameter change will be
immediatelyactive.
NOTE: If a lead is enabled on a nonactivetab of the“Stim”Workspace, the tab will turnthe color
green.Theactivetabstaysthecoloryellow.
NOTE:The lead is disabled and the amplitude is automatically changed to zero when the lead
electrodeconfiguration changes.Theleadelectrode
configurationmustbevalidpriorto activating
thelead.Avalidleadconfigurationmustincludeatleastonepositiveandonenegativeelectrode.