User's Manual

17
LeadsInformationTab[Profile>Leads]
Lead 1 through Lead 4 are the default labels used to
identifytheimplantedleadsin the“Stim”Workspace.Itis
recommended that these names be changed into
something more meaningful, for example the body region
itcovers.
Target Name: for each of the implanted leads,
enterthebodyregion
covered(textfield).
ForeachoftheleadsentertheLotandModelnumber:
Lot #: enter the last four digits of the Lot number
foundontheleadpackaging.
Model#:entertheleadModelnumber.
SystemInformationTab[Profile>System]
Fromthesystemtabthefollowingsystemparameterscanbe
managed:
PeriodicImpedanceinterval:settheperiodicitywith
which you want the system to measure lead
impedance.
Followup Period: A calculated field which displays
the recommended followup time based on the
programmed settings.It is an indicator
ofwhen the
programmerwillrunoutofmemoryandwillbeginto
overwriteolddata.
Ramp Duration: Ramp duration is how long it takes
for the NS to reach the requested amplitude.If set
to8seconds,theNSwilltake8secondstogetfrom0
to
the requested amplitude when a lead is switched
from not enabled to enabled.Ramping also occurs
when the step between the current amplitude and
thenextamplitudeisgreaterthan100mV.
Magnet Turnoff Time: allows you to control how long it takes before a magnet held over the
device
switchesoffdeliveredtherapy.