User's Manual
SpinalModulation,Inc. PhysicianImplantManual
44
AppendixC:ImplantableNeurostimulator
HowSupplied
The Axium™ Implantable Neurostimulator is provided sterile in a double tray assembly and is
intended for single use only.Do not resterilize or reuse, for any reason, because of risk of
infectiontothepatientandmalfunctionofthedevices.
PRECAUTION:DonotresterilizetheINSasitwillcreatearisk ofinfectionor malfunction
ofthedevi ce.
DeviceDescription
TheImplantableNeurostimulatorisafourchannelneurostimulatorthatisonlycompatiblewith
SpinalModulationImplantLeadsandLeadExtensions. Ithasfourportstoallowforstimulation
of up to four leads simultaneously.It has an antenna in the header that allows wireless
communication with the device using the Spinal Modulation Clinical Programmer or Patient
Programmer.
PackageContents
(1)ImplantableNeurostimulator
(3)LeadPortPlugs
(1)TorqueWrench
(1)SterileMagnetSleeve
(1)MedicalAlertCard
(1)ReferencetoImplantManual
DeviceSpecifications
OutputoftheINSisequivalenttotheoutputoftheTNS.Thedeviceisprogrammedincurrent,
impedance is measured by the device, and the appropriate output voltage matches the
impedanceandcurrentprogrammed.
Specifications Range StepSize DefaultValue
PulseAmplitude‐PA(µA)
(Dependingonmeasuredimpedance)
0–6000µA 25µA:0‐2000µA
50µA:2000‐6000µA
0µA
MaximumPulseAmplitude‐Max(µA)
ProgrammablebyPatient
SameasPA SameasPA 0µA
PulseWidth–PW(µs) 40–720µs 10µs 40µs
PulseFrequency‐PF(Hz) 4–100Hz 2Hz 60Hz