PRECAUTION:Donotapplyelectrocauterydirectly totheINS asthiscandamagethe INS
orcauseinterferencewhilecommunicatingwiththeINS.
LeadorExtensionTunneling
TunneltheleadsfromtheanchorsitetotheINSpocket.Whentunnelingtotheabdomen, the
clinicianhastwooptions:
TunneltoamidpointandthencontinuetoINSsite.
Usethe51cmtunnelingtoolandcurvetheconduitaroundthesideofthepatient.
Thefollowingstepsoutlinethesuggestedproceduretotunnelfromtheleadanchorsitetothe
INSpocket:
PRECAUTION:Use extreme care to not damage the lead with the sharp point of the
tunnelingtool.
1. Identifythetunnelingroutebetweentheleadincisionandtheneurostimulatorpocket.
2. Administerlocalanestheticalongthe tunneling route.Additionalsedationmaybe
administeredatthediscretionofthephysician.
3. Bendthetunnelingtoolasnecessarytoconformtothepatient’scontouralongthe
tunnelingroute
4. Withthestrawinplaceonthetunnelingtool,tunnelfromtheINSpockettothelead
anchorsite.
5. Withdrawthetunnelingtoolfromthestraw,leavingthestrawinthesubcutaneous
tunnel.
PRECAUTION:LeadsorExtensionsshouldberoutedadjacenttoone anotherastoprevent
changesinperceivedstimulationfromtheftdetectorsandmetalscreeningdetectors.
6. PasstheendoftheLead(s)orExtension(s)throughthestrawfromtheanchorsitetothe
INSpocketortothemidpointiftunnelingtotheabdomen.Ateachincisionpoint,leave
astrainreliefloopinplaceastominimizethechancesofleadmigration.
7. Removethestrawfromthetunnelbypassingitovertheleads,takingcarenottocause
tractiononthemanddisturbtheleadposition.
ConnectingLeadtoExtension
If an extension is used to connect the Lead to the INS, refer to the section “Percutaneous
ExtensionTunneling”forinstructionsonhowtoconnecttheExtensiontotheLead.