6. Cuttheextensionnear theleadextensionconnector.
7. Discardtheleade xtensionconnector.
8. Preservingsterility,pulltheextensi onoutthroughtheskinexitsite.
9. Discardtheextension.
10. Ifmultipleextensionsareimplanted,repeattheremovalstepsfortheotherextensions.
CreatingtheINSPocketImplant
Oncethelead(s)hasbeenanchored,aNeurostimulatorpocketshouldbecreatedandthelead
tunneledforconnectiontotheINS.
ThefollowingstepsoutlinethesuggestedproceduretocreateanINSpocket:
1. Determine the site for the INS.This should be done before implanting the lead to verify
thereisenoughlengthtoreachtheINSpocketandprovidestrainreliefinthepocket,near
theanchor,andintheepiduralspace.
NOTE: The INS should be located in an area that the patient can easily reach with the
magnetand/orprogrammer:
Intheupperbuttocksalongtheposterioraxillaryline(avoidingthebeltline)
Justovertheabdomenbelowthelowestrib
2. Administerlocalanestheticattheneurostimulator pocketsite.
3. UsebluntdissectiontocreateapocketsothattheINSisparalleltotheskinsurfaceandno
deeperthan2.0cmbelowtheskinsurface.Useelectrocauterytomaintainhemostasis.
4. (Optional)InserttheINSsizertoensurethepocketislargeenoughtoaccommodatethe
INS,allowingextraroomforastrainreliefloopforeachlead.
PRECAUTION:DonotimplanttheINSdeeperthan2.0cm,asthe programmerwillnotbe
abletocommunicatewiththe INS.