todosomayresultindamagetothesheath.Beforereinsertingsheath,verifythereisno
damagetothesheath.
WARNING:If the sheath is not responding to rotation, do notrotate the steering wing
outofplane from thecurve ofthe sheathmo re than90degrees.The tip ofthe sheath
maywhiparoundandcouldcauseharmtothepatient.
12. With the distal end of the
sheath in or at the target foramen, loosen the lead stabilizer,
andadvancetheleadsothatitmovesintotheforamen.Confirmplacementofthe leadon
thedorsalsideoftheforamenusingalateralfluoroscopicview.Verify thattheelectrodes
extendoutoftheshea th.If theelectrodesremainwithinthesheath,stimulation willnot
bepossiblebecauseofhighimpedancereadings.
LooseningtheLeadStabilizer
WARNING:Iftheleadisunabletodeployoutofthesheath,injectsterilewaterorsaline
slowlytoreleasetissuethatmayhaveenteredbetweenthesheathandthe lead.Donot
useexcessivepressurewheninjectingthroughthesheath.