todosomayresultindamagetothesheath.Beforereinsertingsheath,verifythereisno
damagetothesheath.
WARNING:If the sheath is not responding to rotation, do notrotate the steering wing
outofplane from thecurve ofthe sheathmo re than90degrees.The tip ofthe sheath
maywhiparoundandcouldcauseharmtothepatient.
12. With the distal end of the
sheath in or at the target foramen, loosen the lead stabilizer,
andadvancetheleadsothatitmovesintotheforamen.Confirmplacementofthe leadon
thedorsalsideoftheforamenusingalateralfluoroscopicview.Verify thattheelectrodes
extendoutoftheshea th.If theelectrodesremainwithinthesheath,stimulation willnot
bepossiblebecauseofhighimpedancereadings.
LooseningtheLeadStabilizer
WARNING:Iftheleadisunabletodeployoutofthesheath,injectsterilewaterorsaline
slowlytoreleasetissuethatmayhaveenteredbetweenthesheathandthe lead.Donot
useexcessivepressurewheninjectingthroughthesheath.