B. Implanted Lead + Percutaneous Extension Trial A trial is done with an Implant Lead
sutured to the soft tissue just above the spinous process, using the soft tissue anchor to
protect the lead.An extension is tunneled away from the needle insertion site where it
exitstheskin.InasecondprocedureonlytheLeadExtensionisremovedandtheLeadora
new Lead Extension is tunneled to a pocket, the INS is implanted in the pocket, and the
LeadorLeadExtensionisconnectedtotheINS.
PreparingthePatientandDevicesforUse
Leads are designed for placement in the epidural space.Each Lead is accompanied by
accessoriesdesignedtoaidtheclinicianinpositioningthetipoftheLeadnearthetargetDRG.
ToperformaPercutaneousLeadtrial,usethetemporaryTrialLeadKit.
ToperformanImplantLeadtrial,usetheImplantLeadKit.
CAUTION:ThetemporarilyplacedTrialLeadsareintendedforuseforupto30days.Useof
these devices must be performed in accordance with the instructions provided in this
manual.
Usingstandardsteriletechnique,performtheappropriateskinprepping,draping,andinjection
oflocalanesthetictoperformtheepiduralapproachesforpercutaneousleadplacement.
WARNING: The placement of the leads involves some risk, as with any surgical
procedure.Conscious sedation can cause side effects such as systemic toxicity, or
cardiovascular or pulmonary problems.Use caution when sedating the patient.The
patientmustbeawakeandconversantduringportionsoftheproceduretominimizethe
likelihoodofnervedamage.
PlacingtheLead
Leadplacementshouldalwaysbedoneunderfluoroscopicguidance.Theappropriatevertebral
levelforneedleentryshouldbeidentifiedandmarked.
WARNING:Aswithanyspinalepiduralprocedure,potentialrisksofseriousinjuryto the
patient, although extremely rare, include epidural hemorrhage, hematoma, infection,
spinalcordornervecompression,and/orparalysis.
1. Determinethelength oftheleadrequiredtoextendfromthetargetforaminalleveltothe
Neurostimulator implantation site.
Using a 90 cm lead in the lumbar region may cause
difficultyincoilingtheexcesslead.A50cmleadusedintheupperthoracicregionmaynot
belongenoughtoreachtheNeurostimulatorandaleadextensionmaybeneededtobridge
theadditionallength.