The Spex Back Support has been designed to accommodate postural asymmetries and adapt to dierent
morphologies. Once the user has nished using the Spex Back Support, it must be refurbished and
thoroughly cleaned if it is issued to another user. Below are the strict guidelines for refurbishing the Spex
Back Support.
Spex recommends using AHP (Accelerated Hydrogen Peroxide) disinfectant wipes. The disinfection product
must be used as per the manufacturer instructions. Spex takes no responsibility for the eectiveness of the
disinfecting process.
Spex takes no responsibility for contamination, cross infection, or damage.
Re-issuing to another user
Part Procedure
Safety check
All components to be checked for signs of stress or
cracking.
Upholstery covers
Upholstery outer covers must be replaced regardless of
condition.
Backrest shell
Check for signs of stress or cracking. If any stress/cracks
are found the Back Support is to be retired from use and
must be disposed of.
Clean the shell with a mild detergent and quality
disinfectant such as the AHP (Accelerated Hydrogen
Peroxide) disinfectant wipes, ensuring all areas are free
from grime.
Hardware and
components
Ensure the quick release latches and L-bracket mounts are
in place and in safe working order. If any are missing, they
must be ordered. Check for signs of stress or cracking. If
any stress/cracks are found these parts must be retired
from use and replaced immediately.
Inner contouring
section
Ensure each contouring pocket has the correct layers of
foam pads.
Foam section Foam may need replacing.