Instruction Manual
41
9. Stir the probe gently to create a uniform 
sample. Wait about 15 minutes for the 
reading to stabilize.
10. The unit of measure 
will ash on the LCD to 
indicate that the pen is in 
Measurement Mode. When 
the reading is stable, the 
unit of measure will stop 
ashing.
Hold Function
1. Press HLD to freeze the 
current readings on the 
display. “Hold” will appear  
on the LCD.
2. Press HLD again to release 
the hold function. 
Set the temperature coefficient to 
the right value.  The meter is factory 
set to 2.1% per °C and this normally
can provide good results. See page 
9 to set the temperature coefficient
if needed.
Select the normalization temperature. 
The meter is factory set to 25 °C. If 
you need to normalize the meter to 
20 °C, see page 9
Rinse the probe with deionized or 
distilled water before use to remove 
any impurities adhering to the 
electrode. If the meter is not 
used for a long time, please soak the
probe more than 30 minutes to clear 
up the lazy effect of probe.
Dip the probe into the sample. Make 
sure ther
e are no air bubbles trapped 
in the slot of the probe.  To remove 
air bubbles, give the probe a gentle 
stir. Making sure the electrode tip 
is submerged when you stir it.
Stir the probe gently in the sample 
to create a homogenous sample. 
Allow a few seconds to reach temp. 
equilibrium. (Wait about 15 minutes 
can get a stable reading)
  for the details.
 The  unit  of  measurement 
 will  flash   on  the  LCD  to 
 indicate  the  meter   is   in 
 measurement  mode.  When 
 the  reading  is  stable,  the  unit 
 will  stop  flashing. 
8351
4.
5.
6.
7.
8.
9.
5
  “ "
 Press    to  freeze 
  current  readings. The  text
 "Hold"  will  appear  on  the 
 LCD.  P ress  " "  again  to 
 release.
8351
 In  8361  measurement  mode,  press
 " "  to  switch  to  conductivity  or  TDS 
 mode.  The  units  for  conductivity  are
 uS  or  mS.  The  units  for  TDS  are  ppm
 or  ppt. 
Cond.
TDS
6
 In  TDS  measurement  mode,  you  can 
 press  " "  more  than  2  seconds  to 
 select  manual  ranging.  (See  details  in 
 page11)
10.
TDS  Measurement:
Set the  TDS conversion factor to right 
value.  The factory default is 0.50. 
To change the  TDS factors, refer 
Appendix B in page 24.
11.
Salinity  Measurement:
12.
 8371  is  designed  for  salt  measurement.
When in SALT mode, Built-in NaCL 
conductivity to  TDS conversion factor, 
Temp. Coefficient and Normalization 
Temperature are all fixed. 
You may follow up steps 3-10 to 
proceed the salt measurement.
See page 12 
for the detail