Manual
OPERATING PROCEDURE 
Note: For accurate calibration, background noise should 
be less than 10 dB. 
• Determine the sound meter's microphone size. 
• Use the 1/2" Microphone Adapter (2) when 
calibrating a 1/2" diameter microphone. 
• When calibrating a sound meter with a 1" 
microphone, pull off the 1/2" Microphone 
Adapter (2) to use the 1" Microphone Insertion 
Socket (1). 
• Fully insert the sound meter's microphone into the 
insertion socket. 
• Use the 94dB/114dB/Off Slide Switch (4) to turn 
the calibrator on and select either 94 dB or 114 dB. 
• Be sure the sound meter is set to an appropriate 
scale to measure a matching sound of 94 dB or  
114 db. 
• Adjust the sound meter according to it's calibration 
instructions. 
Unfold for Battery Replacement, 
Specifications and Warranty information. 