RH/Temperature SD Card Datalogger 800021 Instruction Manual
RH/Temperature SD Card Datalogger 800021 Copyright ©2010 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval system without the express permission from Sper Scientific. 7720 E. Redfield Rd.
TABLE OF CONTENTS INTRODUCTION. . . . . . . . . . . . . . . 4 FEATURES . . . . . . . . . . . . . . . . . . . 5 MATERIALS SUPPLIED. . . . . . . . . . 7 POWER SUPPLY. . . . . . . . . . . . . . . 8 METER COMPONENTS . . . . . . . . . 9 KEYPAD. . . . . . . . . . . . . . . . . . . . . 11 SETUP. . . . . . . . . . . . . . . . . . . . . . . 12 MEASUREMENT PROCEDURES. . 19 DATALOGGER. . . . . . . . . . . . . . . . 26 BATTERY REPLACEMENT . . . . . . 31 TROUBLESHOOTING. . . . . . . . . . . 32 PC CONNECTION. . . . .
INTRODUCTION The Sper Scientific RH/Temperature SD Card Datalogger (Model 800021) is an extremely accurate meter with fast response times that reads relative humidity (RH), ambient temperature, dew point, wet bulb, and functions as a Type K/J thermocouple thermometer. This meter features a real time SD memory card datalogger. Standard, portable SD memory cards provide unlimited data storage and upload pre-formatted data directly to Excel, eliminating the need for cables or software.
FEATURES • Functions as humidity/ temperature, dew point, wet bulb, and Type K/J thermometer • SD memory card datalogger • Manual datalogger option • Fast response time for humidity • Highly accurate microcomputer circuit • RS232 or USB PC connection • Direct upload of data to Excel • Detachable probe for measuring flexibility • Internal clock and calendar • Touch-tone • Tripod mounting screw • Built-in tabletop stand -5-
FEATURES • Maximum and minimum • Hold function • Auto-power-off • Low battery indicator • Backlight -6-
MATERIALS SUPPLIED • Meter • Probe • SD Card • 6 AA Batteries • Instruction Manual • Soft Carrying Case -7-
POWER SUPPLY This meter can be powered by six AA (1.5 V, UM3) batteries or an optional 9 Volt DC adapter. See page 31 for battery replacement instructions. Plug the adaptor into the power port labeled “DC 9V,” located on the side of the meter. Note) When using the adapter to run the meter, the meter will remain permanently on and the POWER button will be disabled.
METER COMPONENTS Top of Meter Humidity/ Temperature Probe Input Socket Type K/J Thermometer Socket Bottom of Meter SD Card Socket Side of Meter RS232 Output RESET DC 9V Adapter Input Socket -9-
METER COMPONENTS Back of Meter Tripod Mounting Screw Battery Cover Screws Stand Battery Compartment/ Cover Probe Probe Head Probe Handle Probe Plug - 10 -
KEYPAD LCD Display HOLD/ NEXT POWER (ESC, Backlight) REC/ ENTER SET (▲, TIME CHECK) LOGGER (▼, SAMPLING TIME CHECK) - 11 -
SETUP MODE The advanced Setup Mode allows you to customize the following meter preferences and defaults: • Real Time Clock • Decimal Type • Auto Power Off • Touch-Tone • Thermometer Type K/J • Temperature Units • Sampling Time • SD Memory Card Format Note) The setup functions can be performed under any parameter but not while utilizing the datalogger function. Once selections are saved, the meter will default to the selected preferences when the meter is turned off and on again.
SETUP MODE saved, the meter automatically advances to the next setup function (i.e., after setting the real time clock, the meter will automatically enter the decimal type setup function). 1. Press POWER to turn the meter on. 2. Press SET for 2 seconds or longer to enter Setup Mode. 3. Press NEXT to cycle through the setup functions. Note) Press ESC to exit Setup Mode. The meter will return to Normal Mode. Real Time Clock 1. Enter the clock function from Setup Mode (as described above).
SETUP MODE 3. Press ▲ or ▼ to adjust the value. Press ENTER to save the value. 4. Repeat Step 3 to adjust the month, date, hour, minute and second. Note) This procedure adjusts the meter’s internal clock. The internal clock will function when the meter is turned off but only when the batteries have adequate power (not with low battery power.) Decimal Type Although the decimal is commonly expressed as the “.” symbol (such as 20.6 or 1000.
SETUP MODE 2. Press ▲ or ▼ to select Basic (.) or Euro (,). Press ENTER to save the selection. Auto Power Off The meter automatically turns off after 10 minutes of inactivity, however this function can be disabled. 1. Enter the auto power off function from Setup Mode (see page 13). “PoFF” appears on the LCD. 2. Press ▲ or ▼ to select yes (auto power off enabled) or no (auto power off disabled). Press ENTER to save the selection. Touch-Tone 1. Enter the touch-tone function from Setup Mode (see page 13).
SETUP MODE ENTER to save the selection. Thermometer Type K/J 1. Enter the thermometer type function from Setup Mode (see page 13). “tYPE” appears on the LCD. 2. Press ▲ or ▼ to select K or J. Press ENTER to save the selection. Temperature Units 1. Enter the temperature units function from Setup Mode (see page 13). “t-CF” appears on the LCD. 2. Press ▲ or ▼ to select C (degrees Celcius) or F (degrees Farhenheit). Press ENTER to save the selection.
SETUP MODE between successive measurements. To adjust the sampling time (in seconds): 1. Enter the sampling time function from Setup Mode (see page 13). “SP-t” appears on the LCD. 2. Press ▲ or ▼ to adjust the value (1, 2, 5, 10, 30, 60, 120, 300, 600, 800, 1800, 3600 seconds). Press ENTER to save the value. SD Memory Card Format Enabling this function will format the SD memory card to work specifically with your meter. Formatting the SD card will erase any previous memory on the card.
SETUP MODE (format the SD memory card) or no (do not format the SD memory card). Press ENTER to confirm your selection. If selecting yes, “yES Enter” will appear on the LCD, press ENTER again and the meter will format the SD card.
MEASUREMENT PROCEDURES Turning the Unit On/Off 1. Press POWER to turn the meter on. 2. Press and hold POWER for 2 seconds to turn the meter off. Selecting the Function 1. Press and hold FUNC to cycle through the options listed below. Release FUNC when you reach the desired function. • Humidity/temperature “rH” appears on the LCD. • Dew point “dP” appears on the LCD. • Wet bulb “Et” appears on the LCD. • Type K/J thermometer “tP” appears on the LCD.
MEASUREMENT PROCEDURES Note) The meter will default to the last function setting used when it is turned off and on again. Humidity and Temperature 1. Press POWER to turn the meter on. 2. Plug the probe plug into the probe input socket. 3. Press and hold FUNC to cycle through the options until “rH” appears on the LCD. Release the FUNC button. The humidity value (%RH) appears on the upper display and the temperature value (°C or °F) appears on the lower display.
MEASUREMENT PROCEDURES on changing the temperature scale. Dew Point 1. Follow Steps 1-2 in the preceding Humidity and Temperature section. 2. Press and hold FUNC to cycle through the options until “dP” appears on the LCD. Release the FUNC button. The dew point value (°C or °F) appears on the upper display and “dP” appears on the lower display. Wet Bulb 1. Follow Steps 1-2 in the preceding Humidity and Temperature section. 2. Press and hold FUNC to cycle through the options until “Et” appears on the LCD.
MEASUREMENT PROCEDURES lower display. Type K/J Thermometer 1. Press POWER to turn the meter on. 2. Press and hold FUNC to cycle through the options until “tP” appears on the LCD. Release the FUNC button. 3. Plug a thermocouple temperature probe (type K or J) into the type K/J thermometer socket. The LCD will display the measurement value from the temperature probe. 4. “K” or “J” appear on the LCD to indicate the type of thermometer.
MEASUREMENT PROCEDURES Hold Function 1. When measuring any parameter, press HOLD to freeze the reading on the display. “HOLD” will appear on the LCD. 2. Press HOLD again to release the hold function. “HOLD” will disappear from the LCD. Maximum and Minimum To record maximum and minimum readings: 1. When measuring any parameter, press REC to begin recording the maximum and minimum values. “REC” appears on the LCD. 2. Press REC. The maximum value and “REC MAX” will appear on the LCD.
MEASUREMENT PROCEDURES 3. Press REC. The minimum value and “REC MIN” will appear on the LCD. 4. To delete the maximum or minimum value, press HOLD. “REC” appears on the LCD and the meter will begin recording the maximum and minimum values again. 5. To exit the min/max function, press and hold REC for 2 seconds or longer. The meter will return to Normal Mode. Note) The meter cannot be turned off from the memory record function. Exit the function, then press and hold POWER to turn the meter off. Backlight 1.
MEASUREMENT PROCEDURES 2. Press to turn the backlight off. 3. Press to turn the backlight on. View Real Time Clock To view the time function during normal measurement (not during datalogging): 1. Press TIME CHECK. The time information (Year, Month/Date, Hour/Minute) will appear on the lower display of the LCD. View Sampling Time To view the sampling time function during normal measurement (not during datalogging): 1. Press SAMPLING CHECK.
DATALOGGER Preparing the Datalogger 1. Insert the SD card into the SD card socket on the bottom of the meter, ensuring that the front of the SD card faces the back of the meter. 2. Format the SD card as needed (see page 17). 3. Set the real time clock if using the meter for the first time (see page 13). 4. Set the decimal type if using the meter for the first time (see page 14.) Auto Datalogging 1. Set the sampling time to ≥ 1 second. Refer to page 16. 2. Press REC. “REC will appear on the LCD.
DATALOGGER 3. Press LOGGER. “REC” will flash on the LCD and the alarm will sound while the measurement data and time information are saved to memory. 4. To pause datalogging, press LOGGER. The meter will temporarily stop recording and “REC” will stop flashing on the LCD. Press LOGGER again to resume datalogging. “REC” will flash on the LCD. 5. To finish datalogging, while the datalogger is paused, press REC for 2 seconds or longer. “REC” will disappear from the LCD to indicate that datalogging has ended.
DATALOGGER Manual Datalogging 1. Set the sampling time to 0 seconds. Refer to page 16. 2. Press REC. “REC will appear on the LCD. 3. Press LOGGER. “REC” will flash on the LCD and the alarm will sound while the measurement data and time information are saved to memory. The position (location) number will appear on the bottom of the LCD and will also be recorded on the SD card. Note) To enable/disable the touch-tone feature, see page 15. To change the position number, press ▼.
DATALOGGER 4. To finish datalogging, press REC for longer than 2 seconds. “REC” will disappear from the LCD to indicate that datalogging has ended. SD Card Data Structure 1. The first time a SD card is used in this meter, a folder HTA01 will be generated. 2. If the datalogger is being used for the first time, a new file HTA01001.XLS will be generated under the route HTA01\. After exiting the datalogger and executing the function again, the data is saved to the HTA01001.
DATALOGGER 4. The file’s route structure: HTA01\ HTA01001.XLS HTA01002.XLS QQQQQQQ HTA01099.XLS HTA02\ HTA02001.XLS HTA02002.XLS QQQQQQQ HTA02099.XLS HTAXX\ QQQQQQQ QQQQQQQ Note) XX: Maximum value is 10.
BATTERY REPLACEMENT This meter uses six AA (1.5 V, UM3) batteries. When the low battery indicator appears on the LCD, battery replacement is needed. After the icon appears on the LCD, in-spec measurement can still be made for several hours before becoming inaccurate. 1. Press POWER to turn the meter off. 2. Unscrew the battery cover and remove from the meter. 3. Remove the old batteries and replace with six new AA batteries, ensuring correct polarity. 4. Replace the battery cover.
TROUBLESHOOTING System Reset If the meter is not functioning properly (i.e., the system is frozen and the keypad is non-operational), reset the meter: 1. Press POWER to turn the meter on. 2. Use a small tool (such as a disassembled paperclip or a pin) to press the RESET button (located on the right side of the meter under the protective black cover). Wait a few seconds for the meter to restart.
PC CONNECTION To save data from the SD card to a PC (using Excel software): 1. After datalogging is complete, remove the SD card from the meter’s SD card socket. 2. Insert the SD card into the computer’s SD card slot (if built into the computer) or into a SD card adapter (ensuring that the adapter is connected to the computer). 3. Turn the computer on and run the Excel program. 4. Download the saved data file (i.e., HTA01001.XLS, HTA01002.XLS) from the SD card to the computer.
PC CONNECTION channel will upload rH and temperature data; the dP channel will upload dP, rH and temperature data; the Et channel will upload Et, rH and temperature data; and the tP channel will upload tP data only.
PC CONNECTION RS232 PC Serial Interface This meter has a RS232 PC serial interface via a 3.5 mm terminal. The data output is a 16 digit stream that can be utilized for the user’s specific application.
PC CONNECTION DO End word D1 & D8 Display reading, D1 = LSD, D8 = MSD For example: If the display reading is 1234, then D8 to D1 is 00001234 D9 Decimal point (DP) position from right to left 0 = No DP, 1 = 1 DP, 2 = 2 DP, 3 = 3DP D10 Polarity, 0 = Positive, 1 = Negative D11 & D12 Annunciator for Display D13 When send the upper display data = 1 When send the lower display data = 2 D14 4 D15 Start word °C = 01 °F = 02 % RH = 04 - 36 -
PC CONNECTION RS232 Format: 9600, N, 8, 1 Baud rate 9600 Parity No parity Data bit no.
OPTIONAL ACCESSORIES 800060~77 Type K Thermocouple Probes 840057 RS232 Computer Cable 840059 2GB SD Card 840090 Water Resistant Instrument Pouch 840093 Field Tripod 840094 USB Computer Cable 840097 AC Adaptor 850080 Software - 38 -
SPECIFICATIONS General Circuit Custom one-chip of microprocessor LSI circuit Display LCD size: 52 mm x 38 mm Backlight function Measurement Unit Humidity/Temp: %RH/°C or °F Dew point (Humidity): °C or °F Wet bulb (Humidity): °C or °F Type K/J Thermometer: °C or °F Temperature Compensation Automatic temperature compensation for the humidity function and the type K/J thermometer Sampling Time Approximately 1 second of Display Data Output RS232/USB PC computer interface Memory Card SD card 1 GB to 1
SPECIFICATIONS Operating Humidity < 85%RH Power Supply Alkaline or heavy duty DC 1.5 V battery (UM3, AA) x 6 pieces DC 9V adapter input (AC/ DC power adapter is optional) Power Current Normal Operation (without use of the datalogger or backlight): Approximately DC 3.5 mA Datalogger Operation (backlight is off): Approximately DC 28 mA If the backlight is on, the power consumption will increase by approximately 12 mA.
SPECIFICATIONS Humidity Range 5% ~ 95% RH Resolution 0.1% RH Accuracy ≥ 70% RH: ± (3% reading + 1% RH) < 70% RH: ± 3% RH Temperature Range 0°C ~ 50°C (32°F ~ 122°F) Resolution 0.1° Accuracy °C ± 0.8°C °F ± 1.5°F Dew Point (Humidity) °C °F Range -25.3°C ~ 48.9°C Resolution 0.1°C Range -13.5°F ~ 120.1°F Resolution 0.
SPECIFICATIONS Note) The dew point value is automatically calculated from the humidity/ temperature measurement. The dew point accuracy is the sum accuracy value of the humidity and temperature measurement. Wet Bulb (Humidity) °C °F Range -21.6°C ~ 50.0°C Resolution 0.1°C Range -6.9°F ~ 122.0°F Resolution 0.1°F Note) The wet bulb value is automatically calculated from the humidity/ temperature measurement. The wet bulb accuracy is the sum accuracy value of the humidity and temperature measurement.
SPECIFICATIONS Type K/J Thermometer Sensor Type Range Type K -50.0 ~ 0.1°C 1300°C -50.1 ~ -100°C ± (0.4% reading + 0.5°C) ± (0.4% reading + 1°C) -58.0 ~ 2372°F -58.1 ~ -148°F 0.1°F ± (0.4% reading + 1°F) ± (0.4% reading + 1.8°F) -50.0 ~ 0.1°C 1200°C -50.1 ~ -100°C ± (0.4% reading + 0.5°C) ± (0.4% reading + 1°C) 0.1°F ± (0.4% reading + 1°F) ± (0.4% reading + 1.8°F) Type J -58.0 ~ 2192°F -58.
WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of five (5) years from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product.