Product Manual

Spectra Precision® Focus® 8 Series Total Station User Guide 23
Preparation 3
Charging may take up to 5 hours with a completely drained battery which has
been stored for several months without use.
By design Li-Ion batteries should not be charged above 40 °C-45 °C so a blinking
charge light may mean the batteries are too hot for charging. Charging will
resume after the batteries cool down. The charging time will be longer due to
the batteries cool down when charging batteries above 40 °C-45 °C.
If the charge indicator(s) are blinking and the batteries feel cool, it may indicate
a problem with the battery or the charger. If the charge light is still blinking after
trying several batteries which are not warm, there is a problem with the unit or
the batteries themselves.
Conditioning / calibrating a battery
Battery calibration is necessary about once every 6 months or more often if
desired. Calibration insures the reported battery charge remaining is accurate.
Hold down the calibration button on the unit and then insert a battery while
holding the calibration button to begin a battery calibration. Only the battery
which was inserted while the button was depressed will begin calibration.
During a battery calibration the battery will be charged, discharged completely,
and then recharged before finishing. Calibration should complete in roughly 17
hours and the charger vents should not be covered during a calibration cycle.
The blue calibration indicator light(s) will blink slowly (on 1.5 sec, off 2 sec) while
a calibration is in progress and the charge light(s) may be on or off during the
calibration cycle.
When a calibration cycle is completed, the calibration light will stop blinking
remain on until the corresponding battery is removed.
The bottom case temperature may continue to climb up to approximately 43 °C
before temperature regulation is enabled to keep the case from getting warmer.
As the battery voltage drops, the case will cool down and the automatic
temperature limiting will no longer be necessary which minimizes the time it
takes to discharge a battery.
If the case temperature continues to get too hot internally even after
temperature regulation is enabled, there is a secondary fail-safe which will abort
the calibration completely. If an abort occurs, the calibration light(s) will blink
rapidly and battery charging will be re-enabled.