Product Manual

12
6.7 Start Reference Check
Whenworkingduringtemperaturechangesandoverlongdistancestheproductrequiresafrequent
reference check to maintain accuracy and avoid errors caused by drift. The transmitter will do an
automaticuponstartupandafter20minutesofoperation.Itwillrepeatthereferencecheckevery
60minutesandwhenthereisa5°C(9°F)changewithintheproduct.Whencarryingoutworkwhere
accuracy is paramount it is advised to manually prompt a reference check at regular intervals.
PressandreleasetheMENUbuttonattheStandardDisplayandselectReferenceCheck
(Pic23).
Pressing and releasing button 4startstheReferenceCheckconsideringthecurrenttemperatureinside the
housing.Whiletherotorchecksthecorrectpositiontherotationwillbestopped.
Note:Agradevaluehastobeenteredbeforetheunitstartsthereferencecheck.
6.8 Setting Menu
PressandreleasetheMENUbuttonattheStandardDisplayandselectSettings(Pic24).
Press and release button 4toopentheSettingMenu;selectthedesiredfunctionthenpressbutton
4 to open the selected submenu function OR start the selected function.
PleaseseetheSettingMenudetailsattheendoftheuserguide.
6.9 Info
PressandreleasetheMENUbuttonattheStandardDisplayandselectInfo(Pic25).
Buttons6/7canbeusedtotogglebetweenGL,RCandRuntime.
Press and release button 4toconrmtheselection.
TheGL/RCinformation(softwareversion,ID,etc.)ortheruntimeoftheGLwillbedisplayed.
6.10 Service
PressandreleasetheMENUbuttonattheStandardDisplayandselectService(Pic29).Buttons
6/7 can be used to toggle between Calibration Y and Calibration X OR Calibration Z when
set up vertically.
Pic23
Reference
Check
Pic24Settings
Pic25Info
Pic26InfoGL Pic27InfoRC
Pic28Runtimes
Pic29
Service