4. If you want to cancel, press A to select Cancel and then
press C to return to the Adjustments menu.
With the laser pointer on, you can adjust the beam. For more
information, see The Laser Pointer on page 52.
Autofocus calibration
The instrument is equipped with an autofocus function.
Before you can start using the autofocus, the function needs
to be calibrated.
To start the calibration:
1. Level the instrument. The instrument will automatically
check if the compensator is within range before the
calibration is started.
2. Press a to scroll to Autofocus calib and press c.
3. Aim and manually focus at a target at a distance of at least
500 meters, then press c.
C
Set laser pointer
Set mode: Off
Next
Set
>>
-Adjustments-
Tracker collim.
HA/VA collimation
Compensator calib.
Laser pointer
Back...
>>
-Adjustments-
Compensator calib.
HA/VA collimation
Tracker collim.
Laser pointer
Autofocus calib
>>
Aim at Prism and
focus manually
(Dist > 500m)
c