45
c. Press A to scroll to Change face. Then press C
to change face.
d. Aim accurately towards the point.
e. Press C to measure and record angles.
As observations are made on the first face (either face 1 or
face 2), the angle values are stored and the observation
counter increases. When one or more observations has been
taken on each face, and the number of observations on each
face are the same, the software calculates and displays the
new trunnion axis tilt value.
7. Press A to scroll to one of the following:
– Store correction. Then press C to accept the new
trunnion axis tilt value. The Adjustments menu
appears.
– Cancel. Then press C to return to the Adjustments
menu.
Note - The instrument will prohibit a trunnion axis tilt test
if it is made towards a point with an angle less than 15 grads
(13.5 degrees) from the point where the collimation test
was made. The trunnion axis tilt determination accuracy
will improve with a steeper angle towards the measured
point. The minimum distance for the trunnion axis tilt
measurement is 30 m (66 ft.).
Note - If the trunnion axis tilt correction value is greater
than 0.05 grads (0.045 degrees), the message Fail
Remeasure? appears. Press Yes and then repeat the
Trunnion: 0.0003
Cancel
-Current values-
>>
Store correction
C
-Adjustments-
Tracker collim.
HA/VA collimation
Compensator calib.
Laser pointer
Back...
>>