6. Press A to scroll to one of the following:
– New observation. Then press C to continue the
trunnion axis tilt test.
– Change face. Then press C to change face.
– Cancel. Then press C to return to the adjustments
menu.
If you select New observation the number of observations in
both faces appears:
a. Aim accurately in face 2 towards a point at least
15 grads (13.5 degrees) above or below the point where
the collimation test was made at a minimum distance
of 30 m (66 ft.).
b. Press C to measure and record angles.
Aim on target
New observation
Change face
>>
HA: 0.0015
VA: 0.0008
-Current values-
Trunnion coll.
>>
Store correction
C
Face-1 obs: 0
New observation
Change face
Face-2 obs: 0
>>
Face-1 obs: 3
New observation
Change face
Face-2 obs: 3
>>
C